Published
Washington, D.C. : National Aeronautics and Space Administration ; Springfield, Va. : For sale by the National Technical Information Service [distributor], 1972.
Summary
Dynamic response characteristics of six currently used dynamic pressure transducers were investigated by using a large-amplitude sinusoidal-pressure generator. Frequencies between 1 and 15 kilohertz with corresponding peak-to-peak pressure-oscillation amplitudes ranging between 73 and 8 percent of bias pressure and bias pressures between 15 and 300 psia were utilized. Amplitude-ratio data as functions of frequency and pressure level are given for all transducers. The generator design and performance and associated instrumentation are described.